User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Honeywell UDC 3000 Type J øF - High DATE: 15-Feb-96 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 20 NUMBER OF LINES: 153 STANDARD: Thermo Electric MM II (1) STANDARD: Thermo Electric MM II (2) ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON # This procedure is run with the MET/CAL Configuration #1 loaded. 1.001 ASK- P F 1.002 ASK+ X 1.003 DISP Press UUT LOWR DISP button. 1.004 DISP Required Equipment: 1.004 DISP 1.004 DISP Thermo Electric MicroMite II Calibrator; here by refered 1.004 DISP to as the MM II. 1.005 OPBR Is the MM II's DTB number 25-30? 1.006 JMPF 1.009 1.007 STD Thermo Electric MM II (1) 1.008 JMP 1.010 1.009 STD Thermo Electric MM II (2) 1.010 DISP Connect a "J" type connector with "J" type T/C wire 1.010 DISP to UUT as follows: 1.010 DISP ÿ [27][91]1m MMII TO UDC 3000 1.010 DISP ÿ OUTPUT + ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ terminal #17 1.010 DISP ÿ OUTPUT - ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ terminal #18 1.011 HEAD {} 1.012 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.013 HEAD {º ThermoCouple Tests º} 1.014 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.015 HEAD {ÍÍ "J" Type T/C ÍÍ} 1.016 JMP 2.001 1.017 EVAL 2.001 DISP Set the MMII for a "J" type T/C. 2.002 DISP Set the MMII output for 0.0øF. 2.003 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 2.004 ACC 0.0øF 0.5U 2.005 MEME 2.006 MEMC øF 3.2U #! Test Tol 3.2, Sys Tol 0.5, TUR 6.400 (>= 4.00). 3.001 DISP Set the MMII output for 32.0øF. 3.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 3.003 ACC 32.0øF 0.5U 3.004 MEME 3.005 MEMC øF 3.2U #! Test Tol 3.2, Sys Tol 0.5, TUR 6.400 (>= 4.00). 4.001 DISP Set the MMII output for 100.0øF. 4.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 4.003 ACC 100.0øF 0.5U 4.004 MEME 4.005 MEMC øF 3.2U #! Test Tol 3.2, Sys Tol 0.5, TUR 6.400 (>= 4.00). 5.001 DISP Set the MMII output for 200.0øF. 5.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 5.003 ACC 200.0øF 0.5U 5.004 MEME 5.005 MEMC øF 3.2U #! Test Tol 3.2, Sys Tol 0.5, TUR 6.400 (>= 4.00). 6.001 DISP Set the MMII output for 300.0øF. 6.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 6.003 ACC 300.0øF 0.5U 6.004 MEME 6.005 MEMC øF 3.2U #! Test Tol 3.2, Sys Tol 0.5, TUR 6.400 (>= 4.00). 7.001 DISP Set the MMII output for 400.0øF. 7.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 7.003 ACC 400.0øF 0.5U 7.004 MEME 7.005 MEMC øF 3.2U #! Test Tol 3.2, Sys Tol 0.5, TUR 6.400 (>= 4.00). 8.001 DISP Set the MMII output for 500.0øF. 8.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 8.003 ACC 500.0øF 0.5U 8.004 MEME 8.005 MEMC øF 3.2U #! Test Tol 3.2, Sys Tol 0.5, TUR 6.400 (>= 4.00). 9.001 DISP Set the MMII output for 600.0øF. 9.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 9.003 ACC 600.0øF 0.5U 9.004 MEME 9.005 MEMC øF 3.2U #! Test Tol 3.2, Sys Tol 0.5, TUR 6.400 (>= 4.00). 10.001 DISP Set the MMII output for 700.0øF. 10.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 10.003 ACC 700.0øF 0.5U 10.004 MEME 10.005 MEMC øF 3.2U #! Test Tol 3.2, Sys Tol 0.5, TUR 6.400 (>= 4.00). 11.001 DISP Set the MMII output for 800.0øF. 11.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 11.003 ACC 800.0øF 0.5U 11.004 MEME 11.005 MEMC øF 3.2U #! Test Tol 3.2, Sys Tol 0.5, TUR 6.400 (>= 4.00). 12.001 DISP Set the MMII output for 900.0øF. 12.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 12.003 ACC 900.0øF 0.5U 12.004 MEME 12.005 MEMC øF 3.2U #! Test Tol 3.2, Sys Tol 0.5, TUR 6.400 (>= 4.00). 13.001 DISP Set the MMII output for 1000.0øF. 13.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 13.003 ACC 1000.0øF 0.5U 13.004 MEME 13.005 MEMC øF 3.2U #! Test Tol 3.2, Sys Tol 0.5, TUR 6.400 (>= 4.00). 14.001 DISP Set the MMII output for 1100.0øF. 14.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 14.003 ACC 1100.0øF 0.5U 14.004 MEME 14.005 MEMC øF 3.2U #! Test Tol 3.2, Sys Tol 0.5, TUR 6.400 (>= 4.00). 15.001 DISP Set the MMII output for 1200.0øF. 15.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 15.003 ACC 1200.0øF 0.5U 15.004 MEME 15.005 MEMC øF 3.2U #! Test Tol 3.2, Sys Tol 0.5, TUR 6.400 (>= 4.00). 16.001 DISP Set the MMII output for 1300.0øF. 16.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 16.003 ACC 1300.0øF 0.5U 16.004 MEME 16.005 MEMC øF 3.2U #! Test Tol 3.2, Sys Tol 0.5, TUR 6.400 (>= 4.00). 17.001 DISP Set the MMII output for 1400.0øF. 17.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 17.003 ACC 1400.0øF 0.5U 17.004 MEME 17.005 MEMC øF 3.2U #! Test Tol 3.2, Sys Tol 0.5, TUR 6.400 (>= 4.00). 18.001 DISP Set the MMII output for 1500.0øF. 18.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 18.003 ACC 1500.0øF 0.5U 18.004 MEME 18.005 MEMC øF 3.2U #! Test Tol 3.2, Sys Tol 0.5, TUR 6.400 (>= 4.00). 19.001 DISP Set the MMII output for 1600.0øF. 19.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 19.003 ACC 1600.0øF 0.5U 19.004 MEME 19.005 MEMC øF 3.2U #! Test Tol 3.2, Sys Tol 0.5, TUR 6.400 (>= 4.00). 20.001 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.